GET THE APP

Evaluation of Transcutaneous and End-Tidal Carbon Dioxide During
Anesthesia & Clinical Research

Anesthesia & Clinical Research
Open Access

ISSN: 2155-6148

+44 1223 790975

Research Article - (2014) Volume 5, Issue 11

Evaluation of Transcutaneous and End-Tidal Carbon Dioxide During Intravenous Sedation in Volunteers

Kenichi Satoh DDS1*, Hitoshi Miura1, Miho Kumagai2, Masahito Satoh1, Akiyoshi Kuji2 and Shigeharu Joh1
1Division of Dental Anesthesiology, Department of Oral and Maxillofacial Surgery, School of Dentistry, Iwate Medical University, Japan
2Division of Special Care Dentistry, Department of Developmental Oral Health Science, School of Dentistry, Iwate Medical University, Japan
*Corresponding Author: Kenichi Satoh DDS, Division of Dental Anesthesiology, Department of Oral and Maxillofacial Surgery, School of Dentistry, Iwate Medical University, 1-3-27 Chuo-dori, Morioka, Iwate 020-8505, Japan, Tel: +81-19-6515111 Exn. 4331, Fax: +81-19-6520756 Email:

Abstract

Objective: During intravenous sedation, end-tidal carbon dioxide (ETCO2) is usually measured with a nasal cannula or mouth-nose cannula. We compared the measurement accuracy of ETCO2 between these two devices and TC-CO2 and assessed which device is more useful during intravenous sedation in volunteers.

Methods: Eight male volunteers aged 25 to 35 years were evaluated in this single-institution blinded observational trial. After they lay quietly for 5 min without supplemental oxygen, the volunteers received supplemental oxygen by means of each device at a flow rate of 3 L/min for 15 min. Next, midazolam (0.05 mg/kg) was intravenously injected, flumazenil (20 mg) was injected 30 min later, and the ETCO2 and TC-CO2 waveform were recorded.

Results: The differences between ETCO2 and TC-CO2 significantly increased after midazolam injection and decreased after flumazenil injection. The difference between ETCO2 and TC-CO2 using the nasal cannula was greater than that using the mouth-nose cannula. The mean difference between TC-CO2 and ETCO2 ranged from 3 to 9 mmHg after midazolam injection using a nasal mask, and the mean difference ranged from 3 to 6 mmHg after midazolam injection using a mouth-nose cannula.

Conclusions: The difference between ETCO2 and TC-CO2 against TC-CO2 was within the clinically acceptable range. Both the nasal and mouth-nose cannula were useful for ETCO2 measurement with supplemental oxygen by means of each device at flow rate of 3 L/min during intravenous sedation in volunteers.

Keywords: End-tidal CO2; Intravenous sedation; Transcutaneous CO2; Nasal cannula, Mouth-nose cannula

Introduction

Current clinical guidelines recommend capnography as one of the best noninvasive methods with which to assess the adequacy of ventilation in nonintubated patients. The need for capnographic monitoring has dramatically increased with the large number of procedures performed on sedated patients outside the operating room [1].

End-tidal carbon dioxide (ETCO2) waveforms on a capnograph can provide vital information about CO2 retention and respiratory depression [2]. The absence of ETCO2 waveforms indicates a possible state of apnea [3]. ETCO2 measurement is important for identification of intraoperative ventilatory problems in patients undergoing general anesthesia with tracheal intubation [4]. ETCO2 monitoring in nonintubated patients under sedation is frequently necessary in both the intraoperative and postoperative periods. Sedation for medical and dental procedures is a common practice, but it is associated with the risk of excessive respiratory depression [3].

In general, the judgment of whether excessive respiratory depression or apnea has occurred during sedation should involve observation of the presence of breathing, mouth or nose breathing, cyanosis, facial color, movement of the chest wall, and changes in pulse oximetry values. Nevertheless, dental anesthetists performing oral-maxillofacial surgery cannot always immediately judge whether excessive respiratory depression or apnea has occurred because a surgical drape is usually placed over the patient’s face and chest wall during sedation. Pulse oximetry measures the results of inadequate respiration, not inadequate respiration itself, and is therefore not an early indicator of problems [5]. Moreover, supplementary oxygen can delay the detection of respiratory depression by pulse oximetry [6]. The clinical value of ETCO2 monitoring in intravenously sedated patients undergoing dental treatment has not been established.

The two most commonly used standard oxygen delivery devices available for dental treatment are the nasal cannula and mouth-nose cannula. This study was designed to compare the ETCO2 sampling characteristics of these two devices in volunteers with an oxygen (O2) supply. These cannulas deliver O2 through and sample CO2 from both nostrils simultaneously. Whether the delivered O2 significantly dilutes the exhaled gases, thereby providing a falsely low ETCO2 measurement, remains unknown. Arterial blood gas analysis with measurement of the partial pressure of CO2 in arterial blood (PaCO2) is an invasive procedure involving either arterial puncture or placement of an arterial cannula. Arterial blood gas analysis provides only a single measurement. However, a transcutaneous device is used for continuous noninvasive monitoring of PaCO2 [6]. Previous studies have demonstrated a linear correlation between PaCO2 and transcutaneous CO2 (TC-CO2) [7-9]. The results of TC-CO2 monitoring are independent of air leakage [10]. Additionally, this technique has long been validated, is a reproducible measure of the partial pressure of O2 in arterial blood in neonates [11], and is used in adults primarily in the intensive care unit setting [12,13]. Drysdale [14] reported that TC-CO2 monitoring can be used to detect changes in CO2 and is not affected by supplemental O2.

In the present study, the accuracy of ETCO2 measurement using the nasal cannula and mouth-nose cannula were compared with the accuracy of TC-CO2 measurement in volunteer subjects. The aim was to determine which device is more useful during intravenous sedation.

Materials and Methods

Ethics approval

This observational study was approved by the Committee on Clinical Investigation for Human Research (IRB) at Iwate Medical University. Written informed consent was obtained from all volunteers.

Data collection and protocol

Cannula: The nasal cannula used was an Adult CO2 Nasal Sampling Cannula with O2 Administration Cannula® (Novametrix Medical Systems, Inc.; Wallingford, CT). This four-prong nasal cannula delivers O2 via a prong in each nostril and samples exhaled gases via another set of prongs in each nostril.

The mouth-nose cannula used was a Nasal Adapter® (Nihon Kohden; Tokyo, Japan). This cannula is a flow-through capnometer. With this flow-through technology, exhaled CO2 can be monitored by drawing and analyzing an air sample at its source. This cannula has two prongs dedicated to sampling exhaled gases via both nostrils. A twopronged O2 cannula attached to the nasal adaptor in the area between the CO2 prongs and mouth guide delivers O2 to both nostrils.

Capnometry: The CO2 sampling line of the nasal cannula was connected to a sidestream capnometer (Capnomac Ultima; Datex- Engstrom, Helsinki, Finland). The CO2 sampling line of the mouthnose cannula was connected to a flowthrough capnometer (Life Scope BSM-5132; Nihon Kohden).

TC-CO2 monitor: TC-CO2 was measured with a surface monitor (9900MK-II®; Kohken Medical Co., Ltd., Tokyo, Japan). TC-CO2 measurement is based on the principle that a heating element in the electrode elevates the temperature of the underlying tissues. This increases the capillary blood flow and partial pressure of CO2, making the skin permeable to gas diffusion.

Protocol

All volunteers were residents or postgraduate students, and they were not paid to participate in the study. No volunteers had a history of illicit drug use, and all were free from medical conditions such as asthma, respiratory disease, or nasal obstruction. The volunteers comprised eight men ranging in age from 25 to 35 years, with a height of 168.3 ± 2.0 cm and weight of 65.8 ± 8.0 kg. All volunteers were observed with both devices. The observations with the nasal cannula and mouth-nose cannula were performed on different days. The two cannulas were not used at the same time. An electrode was placed on the palmar surface of the forearm of each volunteer and set to a temperature of 43ºC. The sidestream capnometer measured the ETCO2 in the gas expired from the nasal cavity or mouth by means of one of the two devices attached to the nostril or nose. Volunteers received supplemental O2 by means of each device at a flow rate of 3 L/min for 15 min after they lay quietly for 5 min without supplemental O2. This was performed because we usually supply O2 at a flow rate of 2 or 3 L/ min when we use these two devices for dental treatment. Midazolam (0.05 mg/kg) was then intravenously injected, and flumazenil (20 mg) was injected 30 min later. The ETCO2 and TC-CO2 waveforms were simultaneously recorded for 30 min using a PowerLab 16/30T data acquisition system (ADInstruments, Bella Vista, Australia). The measurements were repeated for each device with either nasal or oral breathing. The accuracy of TC-CO2 was compared with that of ETCO2 every minute.

Statistical analysis

Values are presented as mean ± standard error. Statistical analysis was performed using SPSS, version 11.0 (IBM, Chicago, IL, USA). Comparisons between different time points within each group were made using one-way repeated-measures analysis of variance and Dunnett’s multiple-comparison test. The value obtained immediately before supplying the O2 served as the control within each group. Twoway repeated-measures analysis of variance was used to examine between-group differences. Differences were considered statistically significant at a P value of <0.05.

Results

Figure 1 shows the changes in ETCO2 and TC-CO2 on the capnometer waveforms with 3 L/min of O2 supplied using a nasal or mouth-nose cannula during intravenous sedation. The capnometer using the nasal cannula showed a lower peak ETCO2 concentration immediately after midazolam injection than that before injection. The capnometer using the mouth-nose cannula sometimes showed a lower peak ETCO2 concentration immediately after midazolam injection than that before injection. Figure 2 shows the changes in TCCO 2 and ETCO2 using the nasal and mouth-nose cannulas. The TCCO 2 increased after midazolam injection and tended to decrease after flumazenil injection, and there were significant differences between the values 5 to 16 min after injection and the value before injection (P<0.05). Finally, the ETCO2 using the nasal cannula tended to decrease after injection, and that uses the mouth-nose cannula intended to increase immediately after injection. Figure 3 shows the accuracy of TC-CO2 and ETCO2 measurement using the nasal and mouth-nose cannula. The accuracy significantly increased after midazolam injection and decreased after flumazenil injection. The mean accuracy of TC-CO2 and ETCO2 measurement ranged from 3 to 9 mmHg after midazolam injection using a nasal mask, and the mean accuracy ranged from 3 to 6 mmHg after midazolam injection using the mouth-nose cannula.

anesthesia-clinical-research-intravenous-sedation

Figure 1: Changes in TC-CO2 and ETCO2 with 3 L/min of O2 supplied with a nasal or mouth-nose cannula during intravenous sedation. The capnometer in the nasal cannula showed a lower peak ETCO2 concentration immediately after midazolam injection than before injection. The capnometer in the mouth-nose cannula sometimes showed a lower peak ETCO2 concentration immediately after midazolam injection.

anesthesia-clinical-research-nasal-cannula

Figure 2: Changes in TC-CO2 and ETCO2 using a nasal cannula during ntravenous sedation.
The TC-CO2 increased after midazolam injection and tended to decrease after flumazenil injection, and there were significant differences between the values 5 to 16 min after injection and that before injection. The ETCO2 using the nasal cannula tended to decrease after injection, and that using the mouthnose cannula tended to increase immediately after injection. The accuracy increased after midazolam injection and decreased after flumazenil injection. The mean accuracy between TC-CO2 and ETCO2 ranged from 3 to 9 mmHg during sedation using the nasal cannula (n=8).

anesthesia-clinical-research-midazolam-injection

Figure 3: Changes in TC-CO2 and ETCO2 using a mouth-nose cannula during intravenous sedation
The TC-CO2 increased after midazolam injection and tended to decrease after flumazenil injection. The ETCO2 using the nasal cannula tended to decrease after injection, and that uses the mouth-nose cannula intended to increase immediately after injection. The accuracy increased after midazolam injection and decreased after flumazenil injection. The mean accuracy ranged from 3 to 6 mmHg during sedation using the mouth-nose cannula (n=8).

Discussion

In this study, the accuracy of TC-CO2 and ETCO2 measurement using a nasal or mouth-nose cannula was evaluated during intravenous sedation of volunteers. ETCO2 measurement is useful because of its noninvasiveness, continuity, and response time when sudden changes in ventilation occur during intravenous sedation. In the present study, the TC-CO2 continuously increased within 30 min after midazolam injection and decreased after flumazenil injection. Previous studies have reported a linear correlation between PaCO2 and TC-CO2 [6- 8]. Bendjelid [9] reported that evaluation of PaCO2 data on the basis of TC-CO2 values was clinically acceptable, although the relationship between TC-CO2 and PaCO2 varies in patients breathing spontaneously in the intensive care unit. Therefore, when the TC-CO2 increased, the increase in PaCO2 may have occurred after intravenous injection of the sedative drugs. One possible cause of the increased TC-CO2 is that respiratory depression was induced by the intravenous sedative drugs. We are aware of continuous respiratory depression induced by midazolam during intravenous sedation, but have not observed significant increases in TC-CO2.

Clinical guidelines recommend the use of capnography as one of the best noninvasive methods with which to assess the adequacy of ventilation in nonintubated patients [1]. Capnography should be used to monitor ventilation in sedated patients outside the operating room. We generally use pulse oximetry to detect hypoventilation. However, pulse oximetry measures the result of inadequate respiration, not inadequate respiration itself, and is therefore not an indicator of early problems [5]. Supplemental O2 can delay the detection of respiratory depression by pulse oximetry [6]. Additionally, pulse oximeters cannot detect changes in CO2 that may result from reductions in ventilator drive [14]. We mainly use benzodiazepines or propofol for dental anesthesia because intravenous sedation and benzodiazepines inhibit the pathways to the central nervous system by stimulating gammaaminobutyric acid, consequently inhibiting the reduction in the respiratory drive [14]. Capnography enables the measurement of ETCO2 and the early detection of apnea, allowing for immediate intervention to restore ventilation. Moreover, capnographic monitoring with a simple and inexpensive device reportedly reduced the incidence of hypoxemia in patients undergoing colonoscopy during propofol-based sedation [15]. Deitch et al. [16] reported that in adults who underwent propofol sedation on an emergency basis, the addition of capnography to standard monitoring reduced hypoxemia and provided an advance warning for all hypoxemic events. Thus, the need for capnographic monitoring is increasing.

In the present study, the accuracy of ETCO2 and TC-CO2 measurement using a nasal cannula was greater than that using a mouth-nose cannula; however, the data varied depending on whether the subjects breathed spontaneously. The mean accuracy of TC-CO2 and ETCO2 measurement ranged from 3 to 9 mmHg after midazolam injection using a nasal mask, and the mean accuracy ranged from 3 to 6 mmHg during sedation after midazolam injection using a mouth-nose cannula. The mean accuracy of the mouth-nose cannula was lower than that of the nasal cannula because the mouth-nose cannula was designed to measure ETCO2 in expired air during breathing through either the nose or mouth. Nevertheless, the accuracy of the two variables against that of TC-CO2 was within the clinically acceptable range, and the results of this analysis indicate satisfactory agreement of the data between the two devices. The ETCO2 waveform, not the correct ETCO2 value, is required to be able to judge whether excessive respiratory depression has occurred. In this study, both nasal and mouth-nose cannulas were useful for measurement of ETCO2 with the administration of supplemental O2 by means of each device at a flow rate of 3 L/min during intravenous sedation in volunteers.

In conclusion, the accuracy of TC-CO2 and ETCO2 measurement was compared between the nasal and mouth-nose cannula. The accuracy of ETCO2 and TC-CO2 measurement using the nasal cannula was greater than that using the mouth-nose cannula. Nevertheless, the accuracy of the two variables against the TC-CO2 and ETCO2 waveform obtained by means of these two devices was within the clinically acceptable range. Both devices are useful during intravenous sedation in volunteers.

References

  1. Restrepo RD, Nuccio P, Spratt G, Waugh J (2014) Current applications of capnography in non-intubated patients.Expert Rev Respir Med 8: 629-639.
  2. Yanagidate F, Dohi S (2006) Modified nasal cannula for simultaneous oxygen delivery and end-tidal CO2 monitoring during spontaneous breathing.Eur J Anaesthesiol 23: 257-260.
  3. Sakata DJ, Matsubara I, Gopalakrishnan NA, Westenskow DR, White JL, et al. (2009) Flow-through versus sidestreamcapnometry for detection of end tidal carbon dioxide in the sedated patient.J ClinMonitComput 23: 115-122.
  4. Campbell FA, McLeod ME, Bissonnette B, Swartz JS (1994) End-tidal carbon dioxide measurement in infants and children during and after general anaesthesia.Can J Anaesth 41: 107-110.
  5. Miner JR, Heegaard W, Plummer D (2002) End-tidal carbon dioxide monitoring during procedural sedation.AcadEmerg Med 9: 275-280.
  6. Tobias JD (2009) Transcutaneous carbon dioxide monitoring in infants and children.PaediatrAnaesth 19: 434-444.
  7. Griffin J, Terry BE, Burton RK, Ray TL, Keller BP, et al. (2003) Comparison of end-tidal and transcutaneous measures of carbon dioxide during general anaesthesia in severely obese adults.Br J Anaesth 91: 498-501.
  8. Hirabayashi M, Fujiwara C, Ohtani N, Kagawa S, Kamide M (2009) Transcutaneous PCO2 monitors are more accurate than end-tidal PCO2 monitors.J Anesth 23: 198-202.
  9. Bendjelid K, Schütz N, Stotz M, Gerard I, Suter PM, et al. (2005) Transcutaneous PCO2 monitoring in critically ill adults: clinical evaluation of a new sensor.Crit Care Med 33: 2203-2206.
  10. Storre JH, Steurer B, Kabitz HJ, Dreher M, Windisch W (2007) Transcutaneous PCO2 monitoring during initiation of noninvasive ventilation.Chest 132: 1810-1816.
  11. Soto RG, Davis M, Faulkner MJ (2014) A comparison of the incidence of hypercapnea in non-obese and morbidly obese peri-operative patients using the SenTec transcutaneous pCO(2) monitor.J ClinMonitComput 28: 293-298.
  12. Yazawa R, Gama M, Kokawa A, Hirose Y (2008) Are transcutaneous blood gas measurement monitors SenTec Digital Monitor system) useful for the respiratory management of chronic respiratory failure? Masui 57: 453-456.
  13. Rodriguez P, Lellouche F, Aboab J, Buisson CB, Brochard L (2006) Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients.Intensive Care Med 32: 309-312.
  14. Drysdale D (2014) Transcutaneous carbon dioxide monitoring: literature review.Oral Health Dent Manag 13: 453-457.
  15. Friedrich-Rust M, Welte M, Welte C, Albert J, Meckbach Y , et al. (2014) Capnographic monitoring of propofol-based sedation during colonoscopy.Endoscopy 46: 236-244.
  16. Detch K, Meiner J, Chudnofsky CR, Dominich P, Latte D (2010) Dose end tidal CO2 monitoring during emergency department procedural sedation and analgesia with propofol decrease the incidence of hypoxic events? A randomized, control trial. Ann Emerg Med 55: 258-264.
Citation: Kenichi Satoh DDS, Miura H, Kumagai M, Satoh M, Kuji A, et al. (2014) Evaluation of Transcutaneous and End-Tidal Carbon Dioxide During Intravenous Sedation in Volunteers. J Anesth Clin Res 5:460.

Copyright: © 2014 Kenichi Satoh DDS, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Top