Case reportPublished 2016

Ceramic Femoral Head Fracture and Severe Wear of the Cone: A Case Report and Review of the Literature

Authors: Benhamida MK, Benmohamed O, Bekkay MA, Makhlouf H, Bouhdiba S

Journal of Medical Diagnostic Methods

Volume 5 · Issue 1 · Pages 1–4

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

A case of a fractured ceramic femoral head without subsequent trauma, 8 years after a primary total hip arthroplasty, in a 36-year-old man with Osteonecrosis of femoral head (ONFH) is reported. Intraoperative findings were multifragment femoral head without damage of the ceramic insert along with diffuse metallosis and excessive wear of the cone of the stem. Both the stem and the acetabular component were stable. After the removal of ceramic fragments, metallotic tissue excision and careful lavage of the joint and a ceramic femoral head was placed to the existing notched taper of the firmly incorporated stem, the ceramic inlay was kept. At 3-month follow-up examination, the patient had no pain, used no walking aids and had normal activity.

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Longdom does not store it.

Try:

More from this journal

The newest in Journal of Medical Diagnostic Methods.

All its articles

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalJournal of Medical Diagnostic Methods
  • SubjectPathology & Lab Medicine
  • BylineArticles whose byline prints “Benhamida MK”
  • BylineArticles whose byline prints “Benmohamed O”
  • BylineArticles whose byline prints “Bekkay MA”
  • BylineArticles whose byline prints “Makhlouf H”
  • BylineArticles whose byline prints “Bouhdiba S”