Research articlePublished 2018

Determination of Calibration Cycles Using X-ray Tube Output, Workload and Use Factor for Diagnostic X-rays

Authors: Exeter J, Surujpaul PP

Journal of Medical Diagnostic Methods

Volume 7 · Issue 2 · Pages 1–8

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

The stipulated standard time according to annual inspection done under the Health Facility Act 2008 is once per year however, due to varying workloads, use factor, and tube age the calibration cycles can be adjusted. Significance of this work includes determining the optimal calibration time of diagnostic X-ray tubes, resulting in promotion of new knowledge and influencing policies on the need for timely calibrations, additionally it may be used as a marker for urgent calibration in high workload facilities via diagnostic reference per unit to establish an institutional local standard. Parameters involved in this research are X-ray tube output (mGy/mAs), workload (mAmin/ week), use factor and fluence (mGy/cm2). Comparative studies were done at two local hospitals in Guyana namely Georgetown Public Hospital Corporation (GPHC) and Linden Hospital Corporation (LHC). Barium meals (B.M) and intravenous pyelograms (IVP) were measured at Georgetown Public Hospital Corporation. The RaySafe X2 was placed in the path of the X-ray beam at different settings of peak kilovoltage and millamperage- second and the radiation quantities were displayed on the base unit. This research investigated the effect of variation factor on the calibration of the static radiography on a fluoroscopy unit. Factors measured included kerma, tube output, fluence and age of tube output and workload. The differences obtained were significant in the cases of the elevated R/F sensor at 60 cm and at table top of 115 cm and varying peak kilovoltages and milliamperage. The difference in tube output could be attributed to anode surface damage (pitting), age of the tube, filtration, and collimation. A lower workload requires less calibration. In conclusion facilities with a workload of approximately 50 patients do not require more frequent calibration.

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Longdom does not store it.

Try:

More from this journal

The newest in Journal of Medical Diagnostic Methods.

All its articles

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalJournal of Medical Diagnostic Methods
  • SubjectPathology & Lab Medicine
  • BylineArticles whose byline prints “Exeter J”
  • BylineArticles whose byline prints “Surujpaul PP”