Reproducibility and Impact of CT-Scanning on Pulse Wave Velocity Measurement for Cardiovascular Risk Stratification in an Asymptomatic Population

Authors: Thijs L Braber, Niek HJ Prakken, Arend Mosterd, Birgitta K Velthuis

Angiology: Open Access

Volume 4 · Issue 1 · Pages 1–3

Ask about this article

Abstract

Select any words in the title or abstract, or press this and choose a sentence.

Objective: Pulse wave velocity (PWV) measurements for aortic arterial stiffness and coronary CT angiography (CCTA) may help improve cardiovascular risk assessment in asymptomatic people. On the CT-table PWV measurement is an efficient addition to the CT workflow. This study evaluated if on CT-table PWV measurements are influenced by CT anticipation stress and if the PWV measurements are reproducible. Methods: Aortic PWV measurement reproducibility was assessed in 41 asymptomatic male sportsmen (aged 56.5 ± 6.7 years) who underwent CCTA as part of a sports medical evaluation. Three consecutive measurements were performed, two outside the CT-room for intra-observer variability followed by one on the CT-table. Pearson correlation coefficients were assessed for agreement between measurements outside the CT-room and on the CT-table. Bland-Altman analysis of limits of agreement was assessed to evaluate intra-observer variability outside the CT-room. Results: Aortic PWV and systolic blood pressure (SBP) on the CT-table were significantly higher (+0.61 m/s, P = 0.004 and + 7 mmHg, P = 0.003 respectively), with acceptable correlation (Pearson’s correlation 0.8, R² 0.6). The Pearson’s correlation coefficients of PWV measurements outside the CT-room showed good intra-observer agreement, (Pearson’s correlation 0.9, R² 0.8) The mean re-test difference and the 95% limits of agreement outside the CT-room were fair: 0.25 m/s, 95%CI -0.99–1.51 m/s. Conclusion: On the CT-table PWV measurements, although higher due to anticipation stress causing higher SBP, are comparable to off-table measurements at rest. The reproducibility of PWV measurements is good when done prior to a coronary CT-scan and the limits of agreement are acceptable.

Subjects

Cardiology

Ask about this article

Answers come only from this article's title, abstract and keywords, each sentence pointing to where it comes from. Your question is sent to OpenAI to write the answer. Longdom does not store it.

Try:

More from this journal

The newest in Angiology: Open Access.

All its articles

Follow

New articles from what you follow go into your email digest, once you agree to alert emails.

  • JournalAngiology: Open Access
  • SubjectCardiology
  • BylineArticles whose byline prints “Thijs L Braber”
  • BylineArticles whose byline prints “Niek HJ Prakken”
  • BylineArticles whose byline prints “Arend Mosterd”
  • BylineArticles whose byline prints “Birgitta K Velthuis”